The FDA has issued an early alert regarding a potentially high-risk issue involving certain vascular guidewires from Heraeus Medical Components, doing business as Heraeus Medevio.

Heraeus Medevio recommended customers immediately stop using and distributing affected Navilyst Vascular Guidewire, Spectrum Vascular Guidewire and Arrow Marked Guidewire products, according to the FDA’s Nov. 20 alert. Customers were also advised to review inventory and quarantine affected lots.

The issue involves the guidewires’ hydrophilic coating, which can swell inconsistently when hydrated with saline, potentially leading to clumping, flaking, displacement or detachment. Released particles could enter a patient’s bloodstream and lead to a thromboembolic event.

Reported consequences associated with hydrophilic polymer embolization range from asymptomatic deposits to vascular occlusion, ischemia, infarction, hemorrhage, stroke, organ failure, multiorgan failure and death.

As of Nov. 5, Heraeus had not reported any serious injuries or deaths associated with the issue.